Control Standard Endotoxin (CSE) is a widely used standard for endotoxin testing. It is a purified extract
of E. coliO113:H10, the same strain used for the United States Pharmacopeia and the European Pharmacopeia
reference standard endotoxin (RSE).
CSE is used as an economic alternative to the RSE. CSEs are standardized against the RSE as indicated on the
Certificate of Analysis, so that results can be reported in Endotoxin Units (EU) and International Units (IU).
CSE can be used for all LAL testing. A 10 ng/vial CSE is made specifically for use with our Pyrochrome®
chromogenic reagent.
Accessory Product Brochure (PDF)
CSE is used to make standard curves and controls when performing the LAL assay. The concentrations used are dependent on the type of assay (e.g., chromogenic, gel-clot, or turbidimetric) and for photometric methods, the detection range required.
CSE is reconstituted with LRW or equivalent, and is stored at 2–8°C before and after reconstitution. Once reconstituted,
the maximum storage time for the different CSE preparations is listed below.
10 ng/vial: 7 days
0.5 µg/vial: 4 weeks
125 µg/vial: 3 months
CSE should not be frozen.
CSE is reconstituted with LAL Reagent Water (LRW) (or equivalent). Please refer to the Certificate of Analysis when using CSE. A Certificate of Analysis exists for each CSE-lysate lot combination.
The most commonly used size is 0.5 µg/vial. The 0.5 µg/vial is also used for testing depyrogenation ovens. The 125 µg/vial is typically used for special purposes, e.g., making endotoxin challenge articles for validation of depyrogenation processes. The 10 ng/vial CSE is available in packs of five vials.
#E0005-1
Escherichia coli O113:H10, 0.5 µg/vial 1 Pack
#E0005-5
Escherichia coli O113:H10, 0.5 µg/vial 5 Pack
#E0125-1
Escherichia coli O113:H10, 125 µg/vial 1 Pack
#E0125-5
Escherichia coli O113:H10, 125 µg/vial 5 Pack
#EC010-5
Escherichia coli O113:H10, 10 ng/vial,
5 vials per pack, for use with Pyrochrome kits 5 Pack